Set the Micrometer zero setting by touching the Spindle and Anvil using the ratchet. Then press the Reset , Set or Zero Button.
The calibration report should be verified by calibration personnel. Checked by QE SV and QE Department Manager will have to approve the calibration report.
After approval by related Department Manager, the calibration report has to submit to QE Department
Hold the Micrometer with Micrometer Stand and set to Zero again if not Zero.
Micrometer Stand
Micrometer
Piano Wire
Piano Wire attached with
Dial Tension Gauge
Hold the Tension Gauge by wear the finger cot or soft hand glove, Clamp the Piano Wire and measure by touching the Spindle and Anvil to the Piano Wire.
Thickness Gauge
Make sure spin ratchet smooth just one click sound to avoid over turning and to eliminates the clearance
Confirm 5 times at one calibration point and record the reading
Hold the Piano Wire by wear the finger cot, Clamp the Piano Wire and measure by touching the Spindle and Anvil to the Piano Wire.
Make sure spin ratchet smooth just one click sound to avoid over turning and to eliminates the clearance.
Confirm 5 times of calibration reading and record the reading
Reading at 2/3 (1 times)
Reading at 1/3 (2 times)
Reading at 3/3 (2 times)
Hold the Thickness Gauge by wear the finger cot, Clamp the Thickness Gauge and measure by touching the Spindle and Anvil to the Thickness Gauge.
Make sure spin ratchet smooth just one click sound to avoid over turning and to eliminates the clearance
Confirm 5 times at one calibration point and record the reading