Senior Design 2nd Presentation

CR
by Charlotte Rivera0 views·Updated Jul 29, 2013
Powered by AI

Need to present on this topic?

Create your own polished presentation in minutes. Just describe what you need.

Free to try, no credit card required

Transcript

Current imaging techniques cannot effectively locate the stent in relation to the aortic valve.

A modification to the current device to significantly increase its visibility.

Our Solution

How does it work?

From Sound to Sight

Scott Levitt, Andrés Muñoz, Charlotte Rivera, Melissa Stangl

W4

Thank you,

Questions?

Current Treatment Method

Motivation

Limitations of Current Treatment

TAVI device modification and imaging software

Requirement

Echogenic Cover Slip

1 out of 250,000 patients that suffer from severe aortic stenosis

A year long endeavor...

Create a cover slip onto the existing tip of TAVI

  • Have the tip be visible in both fluoroscopy and TEE
  • Act as a constant between both methods by filling in the unknown distance between the tip and valve

Goals and Objectives

TAVI under TEE (ultra sound)

TAVI under Fluoroscopy

  • Finalized device
  • Device meets the specifications
  • Device is echogenic, visible in ultrasound
  • Software system is able to track the device
  • Remain under allocated budget

Specifications

Requirement

Specification

Acoustic Impedance > 1.6-1.7 Rayls

Echogenicity of the Material

Error of Placement with Software

Within 3mm of placement

Image Processing

Tracking the tip from recorded videos and overlay of information

Retain relative dimensions when manufacturing the coverslips

Dimensions/Scalability

Improvement over normal existing tip

Comparison to normal tip

Remain within allocated budget

Cost/Benefit Ratio

Echogenicity of the Material

Nylon

Polyurethane

High Density

Polyethylene

Polycarbonate

Polysulfane

5 biocompatible plastics

Tested the echogenicity of the different tips by submerging them under a water bath and recording ultra sound videos.

Ultra sound video of coverslip made out of Polycarbonate

Graph of Average Pixel Intensity for Each Material

Error of Placement

Cost/Benefit Ratio

Next Fab Grant

BE Course Budget

$1,000

$500

  • The current error of placement for a TAVI is within 3mm.
  • Our aim is to reduce the error to 1mm.
  • This spec will be evaluated on April 23.

Total Project Budget

$1,500

$1,381.68

Total Expenditure

5 manufactured tips

5 plastic materials

2 tubes and 1 plug

1 basic aquarium

cables for video acquisition

Matlab and toolbox

shipping costs

Available Budget

$40.82

$276.33

$30,000

Cost per Tip

Cost of a second TAVI

Evaluation Method

for each Specification

Image Processing

Comparison to normal (existing) tip

  • Existing tip is not available for trial.
  • Qualitative comparison is clear in the videos.

Evaluation of Image Processing

Dimensions/Scalability

  • Image detection software takes ~10.6 seconds to track the tip for a 6 second clip.

Measuring the actual tip's dimensions

Solid works model of the tip

Tips were manufactured 2.5x actual size

A special thanks to...

Future recommendations...

Senior Design Advisers

Howard Herrmann, MD

Robert Wilensky, MD

  • Test for error of placement specification
  • Scaled tips to exact dimensions
  • Production with injection molding
  • Faster processor

Clinical Advisers

Philip Haines, MD

Victor Ferrari, MD

Senior Design Professors

Beth A. Winkelstein, Ph.D

Michael Rizk, Ph.D

Chelsea Salinas, Ph.D